15
SHUTDOWN
30° 120°
.15
.10
.05
CHAMBER
FLOW
MODE CONTROL
REMOTESPAN 1ZERO
DYNACALIBRATOR
MODEL 450
PERMEATION CHAMBER
LOCK
TEMPERATURE °C
PNL TMP
SET
MAIN HEATER
DILUTION
FLOW
CONTROLS
PERMEATION CHAMBER
SPAN 2STANDBY
READ
SPAN 2
FLOW
READ
SPAN 1
FLOW
Figure 7: Model 450 front panel controls and indicators
MODE SELECTOR switch and indicators
The five position switch selects the calibration mode. LEDs indicate the selected
mode.
• STANDBY:Thepermeationchamberoutputisdumpedoutthechambervent,
thedilutionstreamexitsthroughtheoverowvent,andsampleinputisfed
throughtheunitfromthesampleinlettothestreamoutletandontothe
analyzer.
• ZERO:Thepermeationchamberoutputisdumpedoutthechambervent.
Sampleinputisblocked;thedilutionstreamisfedthroughtheSPAN1ow-
meterandoutthestreamoutlettotheanalyzer.
• SPAN1:Withthesampleinputblocked,thechamberoutputandthedilution
streamthroughtheSPAN1owmeteraremixedandfedthroughthestream
outlettotheanalyzerforlowconcentrationcalibration.
• SPAN2:Withthesampleinputblocked,thechamberoutputandthedilution
streamthroughtheSPAN2owmeteraremixedandfedthroughthestream
outlettotheanalyzerforlowconcentrationcalibration.
• REMOTE-UnitdefaultstotheSTANDBYmode.Remotecontactclosures
betweenrear-panelterminalsC,V1,V2,andV3placetheunitinZERO,SPAN1,
orSPAN2mode,respectively.