Valco Instruments Co. Inc.
D3.P65
Rev. 9/07
Printed in USA
Pulsed Discharge
Detector Models
D-3-I-HP and
D-3-I-7890
Instruction Manual
® ®
Europe, Asia, and Africa contact:
tel: + 41 41 925-6200
fax: + 41 41 925-6201
VICI AG International
North America, South America, and Australia/Oceania contact:
tel: 800 367-8424
fax: 713 688-8106
Valco Instruments Co. Inc.
Vista de página 0
1 2 3 4 5 6 ... 24 25

Resumo do Conteúdo

Página 1 - Instruction Manual

Valco Instruments Co. Inc.D3.P65Rev. 9/07Printed in USAPulsed DischargeDetector ModelsD-3-I-HP andD-3-I-7890Instruction Manual® ®Europe, Asia, and Afr

Página 2

6InstallationGeneral Precautions •Do not use plastic/polymer or copper tubes for gas handling andinterconnectons. Use only stainless steel tubing wi

Página 3 - Table of Contents

7Agilent Technologies7890AGC SystemMODEL D3DETECTORFigure 2: A D3 detector in the “Back” position on a 7890ACAPTIVE SCREWS (4)COLUMN INLETELECROMETER

Página 4

8InstallationCAUTION: A D-3-I-HP will not work on a 7890 GC, and aD-3-I-7890 will not work on a 6890. If you are not surewhich model you have, call t

Página 5 - Introduction

9Installation6. Press the <Config> key, then press <Back det> or <Front Det>, corre-sponding to the location where the PDD will be i

Página 6 - Safety Notes and Information

10Installation11. Scroll to select the second line and press <Enter>. The display will read:CAUTION: Instrument power must be turned offand bac

Página 7 - Components of the D3 system

11InstallationGas ConnectionsRemember these three points discussed earlier: (1) all surfaces that contactthe gas stream must be glass or stainless ste

Página 8 - System Requirements

127. Open the cylinder valve to pressurize the regulator once again. Close thevalve and observe the needle of the high pressure gauge for 15 minutes.

Página 9 - Gas Specifications

13Initial Power-UpCAUTION: Always make sure that discharge gas isflowing before heating and powering up the detector.1. Before installing the column,

Página 10 - Installation

14Capillary Column ConnectionIf the capillary column adapter is installed in the column inlet:1. Make a mark on the column 10.3 cm from the end.2. Rem

Página 11

153. Set aside the 1/8" wrench and completely remove the adapter from thecolumn inlet.4. Screw the packed column adapter into the column inlet by

Página 12 - Agilent 7890 GC

This page intentionally left blank for printing purposes

Página 13

16Mode Selection and SetupHelium Ionization ModeSince the PDHID mode provides a better indication of the cleanliness andthe integrity (leak tightness)

Página 14

17Mode Selection and SetupAr-PDPIDChanging the discharge gas from helium to a mixture of 2% argon inhelium changes the photon energy level from the 17

Página 15 - Gas Connections

18TroubleshootingHigh Background CurrentLow SensitivityPurple/bluePinkDisconnect the columnand cap the column inletBaseline current is still high(>

Página 16

19High Noise LevelIf the noise lievel is high:1. See if it improves with the GC fan turned off.2. Check the column insertion position (10.3 cm for cap

Página 17 - Initial Power-Up

20WarrantyThis Limited Warranty gives the Buyer specific legal rights, and a Buyermay also have other rights that vary from state to state. For a peri

Página 18 - Packed Column Connection

21Detector Performance LogIn addition to the occasions indicated in the Comments area of the table below (see Initial Power-Up,page 13), the standing

Página 19 - Testing for Leaks

Table of ContentsIntroductionDescription and Operating Principles ... 1Safety Notes and Infor

Página 20 - Mode Selection and Setup

This page intentionally left blank for printing purposes

Página 21

1IntroductionDescription and Operating PrincipleThe D3 is a non-radioactive pulsed discharge ionization detector (PDID), withmodels available for “plu

Página 22 - Troubleshooting

2Safety Notes and InformationSymbolsATTENTIONRefer to the manual.Installation CategoryThis equipment has been designed for installation category (over

Página 23 - High Noise Level

3Components of the Detector SystemComponents of the detector system are listed in Table 1 below. Check thecontents of the packages to verify that eve

Página 24 - Warranty

4System RequirementsComponents Not Included with the Detector System • Helium (99.999% purity) and other support gases • Ultra high purity grade gas

Página 25 - Detector Performance Log

5IntroductionedoMrotceteDDIHDPDIPDP-rADIPDP-rKDIPDP-eXsagegrahcsiDmuileHeHnirA%2eHnirK%5.1eHnieX%8.0sagreirraCmuileH*****Gas Specifications** Any gas