Setting the Chamber Temperature via Serial Port (RS-232)
Refer to the section titled “Serial Port Communications” on the next page.
Installing the Permeation Device(s)
1. Withthetoolprovided,rotatethepanellockscrew90°counterclockwise.
2. Remove the front panel oven cover to access the chamber cap.
CAUTION: The permeation chamber cap may be warm to
touch.
If the calibrator has been in use, avoid exposure to gas
vapors while opening the chamber cap by using appropriate
mitigation and personal protection equipment.
3. Unscrew the chamber cap with the tool provided.
4. Addorremovethepermeationdevice(s)withthesuppliedforcepsorother
tool appropriate to the job..
5. Secure the chamber cap with the tool provided.
6. Run a pressure/decay test to make sure that the chamber cap is leak tight.
7. Reinstall the front panel assembly and rotate the panel lock screw
90°clockwise.
8. Allowonehourforequilibration.Refertotheseparateinstructionsaccom-
panying the permeation tube to make sure that the selected temperature is
compatible with the permeation tube being used.
The Dynacalibrator is ready to be put into service.
Important Permeation Device Considerations
• Ifmorethanonetubeistobeusedintheovenatthesametime,orderallthe
tubes with permeation rates given at the same temperature.
• Certiedpermeationdevicesshouldbeusedonlyatthetemperaturespecied
on the certificate.
• Usinganypermeationdevicebeyonditsrecommendedtemperaturerange
could result in the destruction of the device by explosion and/or changes
inthemembranecharacteristics.Ifindoubt,contactVICIMetronicsortheir
authorized representative with the part number of the device to determine its
maximum temperature limit.
Initial Power-Up