Valco Instruments Co. Inc.
Rev 9/13
Dynacalibrator
Model 230
Instruction Manual
Valco Instruments Co. Inc.
800 · 367· 8424 sales
713
· 688· 9345 tech
713
· 688· 8106 fax
VICI AG International
Schenkon, Switzerland
Int + 41
· 41 · 925· 6200 phone
Int + 41
· 41 · 925· 6201 fax
North America, South America, and Australia/Oceania contact: Europe, Asia, and Africa contact::
Manufactured by:
Valco Instruments Co. Inc.
8300 Waterbury
Houston, TX 77055
Vista de página 0
1 2 3 4 5 6 ... 41 42

Resumo do Conteúdo

Página 1 - Instruction Manual

Valco Instruments Co. Inc. Rev 9/13 Dynacalibrator Model 230 Instruction Manual Valco Instruments Co. Inc.800 · 367· 8424 sales 713 · 688· 9345 tec

Página 2 - Symbols used in this document

6Concealed DamageConcealed damage is damage which is not apparent until after the equipment has been unpacked and examined or tested. In the event tha

Página 3 - Table of Contents

7Unit LocationIn general, the Dynacalibrator should be as close as possible to the analyzer to be calibrated, as long as:• the ambient temperature is

Página 4 - Table of Contents (continued)

8Rear Panel ConnectionsRemove all shipping caps and shipping plugs from the rear panel bulkhead fittings. All these hardware items should be saved fo

Página 5 - Introduction

9On all units with an internal pump, the supply inlet port is internally connected directly to the suction side of the pump. Since the output of the p

Página 6

10Bypass LoopsOnly those Dynacalibrators with internal pumps are equipped with fittings for bypass loops in either or both the carrier and dilution str

Página 7 - Electronic Flow Control

11MAIN POWER switch, indicator, and circuit breaker The rocker switch provides primary power to all circuits. The LED indicator lights when main powe

Página 8 - Specifications

12CalculationsConcentration of the permeant compound in the span outlet stream is inversely proportional to the carrier flow rate through the chamber.

Página 9 - Getting Started

13Initial Power-UpConnections1. Plug the supplied power cord into the rear mounted power supply. Ensure that the power supply output cable is plugge

Página 10 - Concealed Damage

14Setting the Chamber Temperature ManuallyTo read the current set point: press and hold the button. To read the chassis environment temperature: p

Página 11 - Unit Location

15Setting the Chamber Temperature via Serial Port (RS-232)Refer to the section titled “Serial Port Communications” on page 16.Installing the Permeatio

Página 12 - Rear Panel Connections

AttentionRadiant heat warningEntanglement warningElectrical warningSymbols used in this document

Página 13 - Carrier/Dilution Inlets

16Serial Port CommunicationThe Dynacalibrator Model 230 can be monitored and controlled remotely by means of a serial port connection and a terminal e

Página 14 - Span and Zero Outlets

172. Click “Cancel”. Since no modem is involved, this information is not necessary. The following screen will appear:3. Click “OK”, which reveals

Página 15 - 

186. Select the appropriate COM port and click “OK” to bring up this screen:7. Set the port parameters as indicated above, and click “OK”.8. Now we

Página 16 - Calculations

199. On the dyna230 Properties screen (above), click “Settings” to bring up the screen below. Click “Settings”Click “ASCII Setup...”Serial Port Commu

Página 17 - Initial Power-Up

2010. Click “ASCII Setup” and set the parameters as indicated below. Click “OK”.11. When the screen returns to Dyna230 HyperTerminal (page 18, bot

Página 18

21Analyzer CalibrationAfter the power-up procedure has been com pleted and the chamber temp-erature and permeation device have reached equilibrium, th

Página 19

22Advanced Theory of OperationDynacalibrator operation is based on the principle of mixing a known mass flow of gas derived from a permeation device wi

Página 20 - Serial Port Communication

23this number to indicate the carrier flow (and the orifice used). Refer to page 4 for information about the available flow capacity ranges.Many of the

Página 21

24Configuration 2Pump P1 (Figure 7) draws gas in through the supply inlet on the back panel and delivers it to accumulator AC1. The accumulator smooth

Página 22

25Advanced Theory of Operation

Página 23

Table of ContentsIntroduction General Description...

Página 24 - Entering Commands

26Configuration 5in this configuration (Figure 10), pump P1 draws gas in through the supply inlet on the back panel and delivers it to accumulator AC1.

Página 25 - Shutdown Procedure

27 Electrical DescriptionThe basic wiring of the Model 230 is shown schematically in Figure 11. Dynacali-brators are powered from a benchtop style tr

Página 26 - Advanced Theory of Operation

28MaintenanceSince Dynacalibrators are designed for and generally used in applications that require continuous service, a planned routine maintenance

Página 27 - Configuration 1

2910. Rinse the fan screen and fan filter in clean water and blow dry with low- velocity air. 11. Reinstall the fan, fan filter, fan screen bracke

Página 28 - Configuration 4

30MaintenanceChamber Temperature CalibrationPrior to shipment, the permeation chamber digital temperature control is calibrated against the temperatu

Página 29

31TroubleshootingIf you intend to perform detailed troubleshooting on your Dynacalibrator, please first read the section entitled “Advanced Theory of

Página 30 - Configuration 5

32Troubleshooting4. Isolate the leak by using a small quantity of leak detection fluid on each possible leak point in the suspect loop. Watch carefu

Página 31 - Electrical Description

33Troubleshooting7. Once a section of the loop has been determined to be leak free, go to the next fitting downstream of the point of the previous ch

Página 32 - Maintenance

34

Página 33 - Carrier Flow Rate Verification

35Appendix: Electronic Flow Control OptionTechnical Data for Mass Flow ControllersSpecification Mass Controller DescriptionAccuracy± (0.8% of reading

Página 34

Table of Contents (continued)Electrical Description ...

Página 35 - Troubleshooting

36MAIN POWER switch, indicator, and circuit breaker The rocker switch provides primary power to all circuits. The LED indicator lights when main powe

Página 36 - Component Isolation Method

37Appendix: Electronic Flow Control OptionElectrical SchematicFigure 13: Model 230 electrical schematic, Electronic Flow Control option

Página 37 - Other Pneumatic Issues

38WarrantyThis Limited Warranty gives the Buyer specific legal rights, and a Buyer may also have other rights that vary from state to state. For a per

Página 38 - Model Number Breakdown

1IntroductionGeneral DescriptionVICI Metronics Dynacalibrators® use Dynacal® permeation devices to generate the precise gas concentrations necessary f

Página 39

2IntroductionAs shown in Figure 1a, the carrier stream passes through scrubber #1 and line filter LF1 (30µ) to differential pressure regulator DPR1. T

Página 40 - 

3

Página 41 - Electrical Schematic

4IntroductionSpecificationsRefer to page 35 for additional data on units with the electronic flow control option.

Página 42 - Warranty

5Getting StartedEvery Dynacalibrator is completely calibrated, thoroughly tested and inspected, and carefully packed prior to shipment. The carrier h